August 15, 2017

Multicenter Evaluation of an Image Analysis Device (APAS): Comparison Between Digital Image and Traditional Plate Reading Using Urine Cultures

Background
The application of image analysis technologies for the interpretation of microbiological cultures is evolving rapidly. The primary aim of this study was to establish whether the image analysis system named Automated Plate Assessment System (APAS; LBT Innovations Ltd., Australia) could be applied to screen urine cultures. A secondary aim was to evaluate differences between traditional plate reading (TPR) and the reading of cultures from images, or digital plate reading (DPR).

Methods
A total of 9,224 urine samples submitted for culture to three clinical laboratories, two in Australia and one in the USA, were included in the study. Cultures were prepared on sheep blood and MacConkey agar plates and read by panels of three microbiologists. The plates were then presented to APAS for image capture and analysis, and the images and results were stored for later review.

Results
Image analysis of cultures using APAS produced a diagnostic sensitivity and specificity of 99.0% and 84.5%, respectively. Colonies were detected by APAS on 99.0% of blood agar plates with growth and on 99.5% of MacConkey agar plates. DPR agreed with TPR for colony enumeration on 92.1% of the plates, with a sensitivity of 90.8% and specificity of 92.8% for case designation. However, several differences in the classification of colony morphologies using DPR were identified.

Conclusions
APAS was shown to be a reliable screening system for urine cultures. The study also showed acceptable concordance between DPR and TPR for colony detection, enumeration, and case designation.

Publication: Annals of Laboratory Medicine

Date: January 2017

Authors: Glasson, J., Hill, R., Summerford, M., Olden, D., Papadopoulos, F., Young, S. and Giglio, S.

Citation: Glasson, J., Hill, R., Summerford, M., Olden, D., Papadopoulos, F., Young, S. and Giglio, S., 2017. Multicenter evaluation of an image analysis device (APAS): comparison between digital image and traditional plate reading using urine cultures. Annals of Laboratory Medicine, 37(6), pp.499-504.

doi: 10.3343/alm.2017.37.6.499.